This study evaluated the accuracy of the BD BACTEC FXI Culture System's automated blood volume measurement feature for blood culture testing in simulated pediatric and adult patients. Researchers compared the system's automated measurements against reference measurements obtained through manual weighing, using blood volumes ranging from 2.0 to 15.0 mL for pediatric samples and 40.0 mL for adult samples distributed across culture vials. The automated system demonstrated high accuracy with a mean error of only -0.03 mL per bottle and -0.08 mL per patient, indicating it can reliably measure blood volumes without manual intervention.
Why it matters
Accurate blood volume measurement is critical for detecting bloodstream infections, as insufficient volumes can lead to false-negative results and delayed treatment. This automated system could improve clinical workflow efficiency and ensure appropriate blood collection volumes are achieved, particularly in pediatric settings where volume requirements vary by patient weight and age.
Understand the Science
⚠️ Preprint – Noch nicht peer-reviewed
Dieser Artikel wurde noch nicht von unabhängigen Experten begutachtet. Die Ergebnisse sind vorläufig und sollten mit Vorsicht interpretiert werden.
Objectives: Capturing adequate blood volume for blood cultures is critical for accurate detection of bloodstream infections. Pediatric volume targets vary by age and weight, whereas adult targets are standardized. The BD BACTEC FXI Culture System (FXI) contains an integrated calibrated load cell capable of automatically reporting blood volume measurements for each vial loaded onto the system. This study evaluated the accuracy of the FXI’s blood volume measurements in simulated pediatric and adult patients. Methods: Mock pediatric and adult blood draws were performed, using bagged whole blood, to replicate real-world collection protocols. Syringe-collected blood volumes ranged from 2.0 to 15.0 mL for pediatric patients, depending on mock patient weight, and were fixed at 40.0 mL for adults. Samples were inoculated into BD BACTEC Peds Plus/F, Plus Aerobic/F, and Lytic/10 Anaerobic/F Culture Vials, with a target volume of 2.0 to 10.0 mL per bottle. Reference blood volumes were determined gravimetrically using manually obtained pre- and post-inoculation weights with a blood-specific gravity of 1.055 g/mL and were compared to the automatically measured, gravimetric-based blood volumes reported by the BACTEC FXI Culture System. Results: Automated volume estimates were accurate to a mean error of -0.03 mL per bottle (SD, 0.40 mL; n=168; 95% CI, -0.09 mL, 0.03 mL) and -0.08 mL (SD, 0.79 mL; n=72; 95% CI, -0.26 mL, 0.10 mL) when assessing total volume collected per patient. Conclusions: Our findings demonstrate that the automated system can quantify blood volumes in BACTEC culture vials and support blood volume monitoring for pediatric and adult collections. The gravimetric approach is also amenable to full automation for efficient and accurate blood volume determination.